首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Optimisation of a Flow Injection System with a Piezoelectric Quartz Crystal Detector for the Determination of Inorganic Mercury
Abstract:ABSTRACT

A new method combining flow injection methodology with quartz crystal microbalance (QCM) detection is proposed for inorganic mercury analysis. A modified simplex method is used in order to maximise the observed analytical signals (crystal frequency decrease). Six parameters were optimised: the concentration of the acidic carrier solution, the lengths of the sample loop and mixing coil, and the flow rates of the sample carrier, the reductant stream (SnCl2), and the carrier of the mercury vapour (N2). An increase of 18% in the signal of the centroid of the calibration line was achieved, as well as an increase in the sensitivity from 290 Hz μg? to 313 Hz μg?. Following the proposed procedure, detection limits of 47 μg L? of inorganic mercury were found for sample volumes of 0.5 mL.
Keywords:Quartz crystal microbalance  Piezoelectric crystals  Mercury  Simplex method