首页 | 本学科首页   官方微博 | 高级检索  
     检索      


How Far the Ion-Selectivity Should be Improved; a Task Oriented Limit of Ion-Selective Electrode Design
Abstract:Abstract

A simple formula is proposed for the definition and the calculation of the highest acceptable selectivity coefficient values of ion-selective electrodes. On the basis of the formula which is based on the consideration of the precision of the direct potentiometric measuring techniques one can easily decide whether the selectivity of an electrode is sufficient or not for solving a well defined analytical task correctly. On the other, hand it indicates clearly for ion-selective electrode developing teams or manufacturers how far the ion-selectivity of an electrode should be improved for a certain type of analytical application.

Characteristic selectivity values for a few concrete analytical sample types are given in table together with the selectivity coefficients of the corresponding, existing ion-selective electrodes.
Keywords:ion-selectivity  selectivity coefficient  practical equation for ion-selectivity  ion-selective electrode design