首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Simultaneous Determination of Monoamines,Their Precursor Amino Acids,and Related Metabolites in Mice Brain by High‐Performance Liquid Chromatography with a Coulometric Electrode Array System
Abstract:Abstract

A method based on high performance liquid chromatography with a coulometric electrode array system (HPLC‐coulometric electrode array) using C18 column has been developed for the simultaneous determination of norepinephrine (NE), epinephrine (E), l‐3,4‐dihydroxyphenylalanine (l‐DOPA), p‐tyrosine (p‐TYR), dopamine (DA), m‐tyrosine (m‐TYR), 5‐hydroxytryptamine (5‐HT), homovanillic acid (HVA), and 5‐hydroxyindoleacetic acid (5‐HIAA) in mice brain. The chromatography was performed using a C18 column (250 mm×4.6 mm i.d. and 5 µm) with sodium acetate buffer (pH 5.0, 0.05 M) and methanol as the mobile phase. Elution of analytes was carried out at a flow rate of 1.0 ml/min. The nine compounds were monitored using an ESA electrochemical detector. Potentials of three electrodes in series were set at 200, 500, and 700 mV, respectively. Optimization of the pH of the mobile phase and the proportion of methanol were also considered. The minimal detection limits were 2–8 ng/ml. Linear (r=0.99) detector performances were observed within a range of 10~2000 ng/ml. Recoveries for the nine compounds in spiked samples were over 90% and the relative standard deviations (RSD) were less than 4.0%.
Keywords:HPLC  electrochemical detector  coulometric array  monoamine  mice brain
设为首页 | 免责声明 | 关于勤云 | 加入收藏