首页 | 本学科首页   官方微博 | 高级检索  
     


Simultaneous Determination of Gallium(III) and Indium(III) in Urine and Water Samples with Cloud Point Extraction and by Inductively Coupled Plasma Optical Emission Spectrometry
Abstract:A simple, sensitive, and selective method for the determination of gallium and indium in different samples at trace levels is presented. This method was based on complexation of analyzes with 2-(5-bromo-2- pyridylazo)-5-diethylaminophenol (5-Br-PADAP) in the presence of t-octylphenoxy-polyethoxyethanol (Triton X-100). After phase separation, the analyzed concentrations were determined by inductively coupled plasma optical emission spectrometry. Quantitative extraction of gallium and indium was performed at pH 7.0, 1.7 mmol L?1, 5-Br-PADAP, 1.3% (w/v) Triton X-100 and at 75°C. The relative standard deviations (RSD) of this method were between 0.3% and 1.6% (C = 20 ng mL?1, n = 9). The calibration curve is linear over the concentration range 6–200 ng mL?1 for gallium and 2–200 ng mL?1 for indium, respectively. The limits of detection (LOD) for gallium and indium were 0.72 and 0.28 ng mL?1, respectively. The results showed the developed method was not susceptible to matrix effects, providing recoveries between 98% and 102%. The accuracy of the developed method was evaluated by the analysis of spiked certified reference materials. The developed method was successfully applied to gallium and indium determination in urine and lake water with satisfactory results.
Keywords:Cloud point extraction  Gallium  Indium  Urines and lake water samples
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号