首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of Cadmium and Lead in Sea Water after Extraction Using Electrothermal Atomisation. Minimisation of Interferences from Co-Extracted Sea Salts
Abstract:Abstract

Interferences from small amounts of sea salt in the determination of cadmium and lead were investigated. Test samples were made to mimic solutions obtained after extraction of sea water for trace metal analysis. Sea salt concentrations in the range 0–400 mg/l were investigated. Background absorption from this salt was high but easily minimised through the addition of nitric acid (2 % v/v). Sensitivity reductions due to the salt were considerable and varied from one graphite tube to the other. This problem was overcome by the addition of 1000 mg lanthanum per litre of sample. The positive effect of the lanthanum matrix modification is thought to be due to a change of the graphite tube surface.
Keywords:Cadmium  lead  sea water  extraction  electrothermal atomisation  atomic absorption  matrix modification  lanthanum.
设为首页 | 免责声明 | 关于勤云 | 加入收藏