首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development and Application of an Immobilized Enzyme Electrode for the Determination of Sulfite in Foods and Feeds
Abstract:Abstract

An enzyme electrode was developed using the enzyme sulfite oxidase (EC 1.8.3.1) immobilized onto pig intestine. All experimental parameters such as pH, temperature, buffer constituent concentration, etc., were thoroughly investigated and optimized when appropriate. Hydrogen peroxide was monitored amperometrically. The response to sulfite concentration was linear in the range of 5.2 × 10?6 to 1.0×10?3M. The proposed method was found to have a correlation coefficient of 0.952 when evaluated versus the standard titration method.
Keywords:Enzyme Electrode  Sulfite Determination  Sulfite Oxidase  Foods &  feeds