首页 | 本学科首页   官方微博 | 高级检索  
     


Investigations on Bioanalytical Chemistry Part I. On Differential Pulse Voltammetry of Bilirubin
Abstract:Abstract

The electrochemical behaviour of the bilirubin in many kinds of supporting electrolytes on a glassy carbon electrode (GCE) and a hanging mercury drop electrode (HMDE) was investigated by means of anodic or cathodic differential pulse voltammetry. The influences of different methods of pre-treatment of the glassy carbon electrode was also discussed. In Na2B4.O7-KH2PO4 buffer solution, the linear range was 2×10?9-1×10?9 mol/l and the detection limit was 3.3×10?9 mol/l by anodic differential pulse voltammetry at GCE. A linear relationship holds between the peak current and the concentration of bilirubin in a concentration range of 1×10?9-4×10?7 mol/l with good precision and accuracy, and the limit of detection was 2×10?10 mol/l, when cathodic differential pulse adsorption voltammetry at HMDE was used.
Keywords:Bilirubin  trace analysis  bioanalytical chemistry  stripping analysis  voltammetry
设为首页 | 免责声明 | 关于勤云 | 加入收藏