Authors:Conley Justin M  Symes Steven J  Kindelberger Scott A  Richards Sean M
Institution:Department of Biological and Environmental Sciences, University of Tennessee at Chattanooga, 615 McCallie Avenue, Chattanooga, TN 37403, USA. justinmconley@hotmail.com
Abstract:Herein, a new method for the detection of 13 different pharmaceuticals and one metabolite in surface water at low ng/L levels is described. The method utilizes ultra performance liquid chromatography-tandem mass spectrometry and a solid-phase extraction sample preparation. Mean method detection limits were low (4.10 ng/L) and overall solid-phase extraction recovery and reproducibility was adequate (mean recovery, 77.9%; mean RSD, 7.3%). The method allows for quick run times and minimal solvent use as compared with other previously reported high performance liquid chromatography-based methods. Application of this method for the detection of pharmaceuticals in Tennessee River surface water determined that caffeine, sulfamethoxazole, and carbamazepine were frequently detected (100% of samples). Trimethoprim was moderately detected (30% of samples); acetaminophen, atorvastatin, and lovastatin were infrequently detected (10% of samples); and ciprofloxacin, diltiazem, fluoxetine, levofloxacin, norfluoxetine, ranitidine, and sertraline were not detected. This study reports the first detection of lovastatin in surface water.