Enzymatic formation of ions and their detection at a three-phase electrode
Authors:Rubin Gulaboski  Carlos M. Pereira  M. Natalia D. S. Cordeiro  Ivan Bogeski  Fernando Silva
Affiliation:(1) Departamento de Quimica, Faculdade do Ciências, Universidade do Porto, 4169-007 Porto, Portugal;(2) Institute of Physiology, University of Saarland, Homburg, Germany
Abstract:An electrochemical method for the detection of enzymatically created anions is described that uses a thin-film electrode with decamethylferrocene as an electroactive redox probe. The enzymatic oxidation of glucose with enzyme glucose oxidase produces gluconic acid as a final product. The oxidation of decamethylferrocene dissolved in the thin-nitrobenzene film, that is spread on the working graphite electrode and submerged in the aqueous solution containing glucose and glucose oxidase, is followed by the up-take of gluconate anions from the aqueous phase to nitrobenzene. The peak currents of the square-wave voltammetric responses of that system are a linear function of the glucose concentration in the milimolar range from 0.1 mmol/L to 0.7 mmol/L (R2=0.994).
Keywords:Glucose  Enzymatic reactions  Biosensor  Thin-film electrode  Ion transfer  Decamethylferrocene
本文献已被 SpringerLink 等数据库收录!