首页 | 本学科首页   官方微博 | 高级检索  
     


On-line preconcentration of trace copper for flame atomic absorption spectrometry using a spherical cellulose sorbent with chemically bound quinolin-8-ol
Authors:Ernest Beinrohr  Miroslav Čakrt  Ján Garaj  Miroslav Rapta
Affiliation:Department of Analytical Chemistry, Slovak Technical University, Radlinskeho 9, CS-812 37 Bratislava Czechoslovakia
Abstract:Preconcentration was effected using a 50 mm × 2 mm i.d. minicolumn packed with a spherical cellulose sorbent with chemically bound quinolin-8-ol function groups (Ostorb Oxin). The column was connected to the nebulizer of the atomic absorption spectrometer and the sample solution and eluent (2 M hydrochloric acid) were sucked through it at a flow-rate of 2–3 ml min?1 by utilizing the negative pressure of the nebulizer. The experimental design was tested with the determination of traces of copper. Peak-area and peak-height measurements were compared. Owing to the simple calibration, the former method was used for quantification. The dynamic range was from 0.3 ng ml?1 (detection limit) to 5 μg ml?1 (breakthrough). The reproducibility in the concentration range 25 ng ml?1-5 μg ml?1 was better than 5%. Water-soluble inorganic salts, ammonia and sodium hydroxide were analysed. The accuracy of the results was checked by anodic-stripping voltammetry and by electrothermal AAS.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏