Accurate measurement of ultralow loss in a high-finesse Fabry-Perot interferometer using the frequency response functions
Authors:N Uehara  K Ueda
Institution:(1) Institute for Laser Science, University of Electro-Communications, 1-5-1 Chofugaoka, Chofu, 182 Tokyo, Japan
Abstract:We describe the accurate measurement of ultralow loss in a high-finesse Fabry-Perot interferometer using a diode-pumped Nd:YAG laser locked to the longitudinal mode with an active frequency-stabilization technique. By measuring the resonance full width and the free spectral range with the frequency response functions, and by measuring the transmission efficiency on resonance, the finesse and the loss at 1064 nm are accurately measured to be 78100 ± 1200 (reflectance of 99.99598 ± 0.00006%) and 1 5.9 ±2.0 × 10–6 (15.9 ± 2.0 ppm), respectively.
Keywords:06  00  07  60  42  60
本文献已被 SpringerLink 等数据库收录!