首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Measurement of240Pu/239Pu ratio by fission track method and inductively coupled plasma mass spectrometry
Authors:C K Kim  Y Oura  Y Takaku  H Nitta  Y Igarashi  N Ikeda
Institution:(1) Department of Chemistry, University of Tsukuba, 305 Tsukuba, Ibaraki, Japan;(2) Marubun Co., Odenma-cho 8-1, Nihonbashi, Chuo-Ku, 103 Tokyo, Japan;(3) Division of Radioecology, National Institute of Radiological Sciences, Isozaki 3601, 311-12 Nakaminato, Ibaraki, Japan
Abstract:Fission track /FT/ method and inductively coupled plasma mass spectrometry /ICP-MS/, that is a new analytical technique for the analysis of trace element, were used for the measurement of240Pu/239Pu ratios in environmental samples. The results obtained by both methods are in agreement within the relative deviation of 9–13%. The precision in ICP-MS was found to be better than in it the FT-method. These methods are applicable to measure the Pu isotopes ratio at low concentration levels in environmental samples.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏