Preconcentration of trace heavy metals in water by coprecipitation with magnesium oxinate for electrothermal atomic absorption spectrometry
Authors:Zuo -Sheng Chen  Masataka Hiraide  Hiroshi Kawaguchi
Institution:(1) Department of Materials Science and Engineering, Nagoya University, Chikusa-ku, 464 Nagoya, Japan
Abstract:Nanogram quantities of heavy metals in 200 ml of water were quantitatively preconcentrated by coprecipitation with magnesium oxinate at pH 9–9.5. The precipitate was collected on a 1-mgrm Nuclepore membrane filter and then dissolved in 5 ml of 1 mol/l nitric acid. After diluting to 10 ml with water, a 10- or 20-mgrl aliquot of the solution was directly analyzed by electrothermal atomic absorption spectrometry using deuterium background correction. The background absorption due to the collector magnesium oxinate was negligibly small. The optimized coprecipitation technique has been applied to the determination of cobalt, nickel, copper, cadmium and lead at the ppt level in river and seawater. The relative standard deviations were within 10% and the detection limits were 0.025 (for Co), 0.019 (for Ni), 0.006 (for Cu), 0.001 (for Cd) and 0.017 mgrg/l (for Pb). Blanks through the whole procedure were not detectable.
Keywords:preconcentration  coprecipitation  magnesium oxinate  trace heavy metals  water analysis  electrothermal atomic absorption spectrometry