Chemical modifiers for direct determination of cobalt in coal combustion residues by ultrasonic slurry-sampling-ETAAS
Authors:M Felipe-Sotelo  A Carlosena  E Fernández  P López-Mahía  S Muniategui  D Prada
Institution:Department of Analytical Chemistry, University of La Coru?a, Spain.
Abstract:Five modifiers were tested for the direct determination of cobalt in coal fly ash and slag by ultrasonic slurry-sampling electrothermal atomic absorption spectrometry (USS-ETAAS).The furnace temperature programs and the appropriate amount for each modifier were optimized to get the highest signal and the best separation between the atomic and background signals. Nitric acid (0.5% v/v) was the most adequate chemical modifier for cobalt determination, selecting 1,450 degrees C and 2,100 degrees C as pyrolysis and atomization temperatures, respectively. This modifier also acts as liquid medium for the slurry simplifying the procedure. The remaining modifiers enhanced the background signal, totally overlapped with cobalt peak. The method optimized gave a limit of detection of 0.36 microg g(-1), a characteristic mass of 13 +/- 1 pg and an overall-method precision which is highly satisfactory (<7%, RSD). The method was validated by analyzing two certified coal fly ash materials, and satisfactory recoveries were obtained (83-90%) and no statistical differences were observed between the experimental and the certified cobalt concentrations. Additionally, certified sediment, soil and urban particulate matter were assayed; again good results were obtained. The developed methodology was used to determine cobalt in several coal combustion residues from five Spanish power plants.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号