首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of a radioimmunoassay of alpha 1-microglobulin (alpha 1-m) with simplified procedures
Authors:K Matsui  H Moriuma  C Mishima  M Honda  M Tomonobu  K Kanao  H Fushimi
Institution:Department of Nuclear Medicine, Sumitomo Hospital, Osaka, Japan.
Abstract:A newly established double antibody radioimmunoassay (RIA) was fundamentally and clinically evaluated. Original procedures were partially modified as follows: Sample volume for serum and urine was changed to 25 microliters, and thus 200 mg/l of alpha 1-m standard was prepared using 50 microliters of original standard solution (100 mg/l). The results were satisfactory in sensitivity (0.3 mg/l obtained from -2SD method), intra-assay precision with its coefficient variation (CV) ranging from 3.0 to 7.4%, interassay precision with its CV ranging from 3.0 to 10.7%, and recovery with the mean value of 102.4% in serum and 108.2% in urine respectively. There were no changes about alpha 1-m value between diluted (2 times) and undiluted with high concentration samples. Normal levels of alpha 1-m were less than 25 mg/l in serum and less than 10 mg/l in urine. The present results indicate that the determination of alpha 1-m could be very simple and useful for the most sensitive screening test for the evaluation of renal function.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏