DESIGN AND PERFORMANCE OF THE OKAZAKI LARGE SPECTROGRAPH FOR PHOTOBIOLOGICAL RESEARCH
Authors:Masakatsu,WatanabeMasaki,Furuya&dagger  Yasuhiro,Miyoshi&dagger  Yasunori,Inoue&dagger  Isao,Iwahashi&Dagger  Koichi,Matsumoto&Dagger  
Affiliation:National Institute for Basic Biology, Okazaki National Research Institutes, Okazaki, Aichi 444;Biology Department, Faculty of Science, University of Tokyo, Hongo, Tokyo 113.;Naka Works, Hitachi Ltd., Katsuta, Ibaraki 312, Japan
Abstract:A computer-operated spectrograph was recently built at Okazaki, Japan. Different specimens can be placed on a horseshoe-shaped focal curve (10 m long) covering a wavelength range of 250 to 1000 nm so they can be irradiated simultaneously. The linear dispersion is about 0.8 nm/cm. The photon fluence rate on the focal curve is 5 x 1015. photons x cm-2 x s-1 at 300nm and 1 x 1016 photons x cm-2 x s-1 at 600 and at 900 nm. The spectral half width is 5.5 nm or less on the focal curve. The stray light content is about 10-5 of the main peak at the peak wavelength ± 100 nm. Specimens are set in microcomputer-controlled threshold boxes so that wavelengths, photon fluence rates, photon fluences and timing of irradiations are controlled automatically according to a pre-programmed schedule. An optical fiber system is also provided for remote irradiations.