Identification and quantification of bisphenol A and bisphenol B in polyvinylchloride and polycarbonate medical devices by gas chromatography with mass spectrometry
Authors:Pascal Gimeno,Coline Spinau,Nelly Lassu,Annie‐Franç  oise Maggio,Charlotte Brenier,Laurent Lempereur
Abstract:A gas chromatography with mass spectrometry method has already been developed and published for the identification and quantification of 14 phthalates and five nonphthalate plasticizers in polyvinylchloride medical devices. In order to assay, in addition to plasticizers, bisphenols A and B possibly present in polyvinylchloride samples, this previous method was extended to the assay of these additional potential endocrine disruptors. Furthermore, as bisphenol A could also be present in polycarbonate samples, the method used for the polyvinylchloride sample was tested and validated for the assay of bisphenols A and B in polycarbonate medical devices. The separation of all compounds, including bisphenols A and B, is obtained on a cross‐linked 5%‐phenyl/95%‐dimethylpolysiloxane capillary column using a temperature gradient. For both plastics, samples are dissolved in tetrahydrofuran followed by a precipitation of the plastic by addition of ethanol. Results obtained point out residual bisphenol A amounts for polycarbonate samples ranging from 0.6 to 0.8% and for polyvinylchloride samples less or equal to 5 ppm. No bisphenol B was detected in the samples tested. For bisphenols A and B, mean recoveries obtained on spiked polyvinylchloride or polycarbonate sample preparations ranged from 87 to 108% in accordance with in‐house specification (80–110%).
Keywords:Bisphenols  Gas chromatography with mass spectrometry  Medical devices  Phthalates  Plasticizers
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号