首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of phenols by HPLC down to PPT levels
Authors:D A Baldwin  J K Debowski
Institution:(1) Division of Processing and Chemical Manufacturing Technology, Council for Scientific and Industrial Research, P.O. Box 395, 0001 Pretoria, Republic of South Africa
Abstract:Summary A mixture of eleven priority pollutant phenols (PPP) have been separated using a Dionex 4000 chromatograph; VAL-U-PAK ODS (250 × 4.6 mm I.D.) column; eluents A: CH3CN/0.2M NaH2PO4/CH3COOH 40/60/1 and B: CH3CN with/without gradient; detectors: UV-254 nm and Pulse Amperometric Detector (PAD) with glassycarbon electrode (+1.20V). With preconcentration unit (guard RP-18 column) on line and PAD, detectability limit at ten ppt was achieved. The method has been applied for determination of PPP in tap-water and in leachates from sanitary and industrial landfill sites.
Keywords:Priority pollutant phenols separation  Column liquid chromatography reversed phase  On line preconcentration  Pulse amperometric detector  Detectability at ppt level
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏