Determination of phenols by HPLC down to PPT levels |
| |
Authors: | D A Baldwin J K Debowski |
| |
Institution: | (1) Division of Processing and Chemical Manufacturing Technology, Council for Scientific and Industrial Research, P.O. Box 395, 0001 Pretoria, Republic of South Africa |
| |
Abstract: | Summary A mixture of eleven priority pollutant phenols (PPP) have been separated using a Dionex 4000 chromatograph; VAL-U-PAK ODS (250 × 4.6 mm I.D.) column; eluents A: CH3CN/0.2M NaH2PO4/CH3COOH 40/60/1 and B: CH3CN with/without gradient; detectors: UV-254 nm and Pulse Amperometric Detector (PAD) with glassycarbon electrode (+1.20V). With preconcentration unit (guard RP-18 column) on line and PAD, detectability limit at ten ppt was achieved. The method has been applied for determination of PPP in tap-water and in leachates from sanitary and industrial landfill sites. |
| |
Keywords: | Priority pollutant phenols separation Column liquid chromatography reversed phase On line preconcentration Pulse amperometric detector Detectability at ppt level |
本文献已被 SpringerLink 等数据库收录! |