首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Preliminary results on the determination of ultratrace amounts of cadmium in tea samples using a flow injection on-line solid phase extraction separation and preconcentration technique to couple with a sequential injection hydride generation atomic fluorescence spectrometry
Authors:Duan Taicheng  Song Xuejie  Jin Dan  Li Hongfei  Xu Jingwei  Chen Hangting
Institution:

aState Key Laboratory of Electroanalytical Chemistry, National Research Center of Electroanalytical and Spectroanalytical Chemistry, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun 130022, China

bFaculty of Chemistry, Jilin Universtity, Changchun 130026, China

cLaboratory of Rare Earth Chemistry and Physics, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun 130022,China

Abstract:In this work, a method was developed for determination of ultra-trace levels of Cd in tea samples by atomic fluorescence spectrometry (AFS). A flow injection solid phase extraction (FI-SPE) separation and preconcentration technique, to on-line couple with a sequential injection hydride generation (SI-HG) technique is employed in this study. Cd was preconcentrated on the SPE column, which was made from a neutral extractant named Cyanex 923, while other matrix ions or interfering ions were completely or mostly separated off. Conditions for the SPE separation and preconcentration, as well as conditions for the HG technique, were studied. Due to the separation of interfering elements, Cd hydride generation efficiency could be greatly enhanced with the sole presence of Co2+ with a concentration of 200 μg L−1, which is much lower than those in other works previously reported. Interferences on both the Cd separation and preconcentration, and Cd hydride generation (HG) were investigated; it showed that both the separation and preconcentration system, and the HG system had a strong anti-interference ability. The SPE column could be repeatedly used at least 400 times, a R.S.D. of 0.97% was obtained for 6 measurements of Cd with 0.2 μg L−1 and a correlation coefficiency of 1.0000 was obtained for the measurement of a series of solutions with Cd concentrations from 0.1 to 2 μg L−1. The method has a low detection limit of 10.8 ng L−1 for a 25 mL solution and was successfully validated by using two tea standard reference materials (GBW08513 and GBW07605).
Keywords:Cd  Tea sample  AFS  On-line preconcentration and separation  Hydride generation  Cyanex 923
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏