Preconcentration of inorganic selenium species (Se (IV) and Se (VI)) in an alumina filled microcolumn and on-line determination by hydride generation atomic absorption spectrometry
Authors:A. Larraya   M. G. Cobo-Fernández   M. A. Palacios  C. Cámara
Affiliation:(1) Departamento de Química Analítica, Facultad de C. Químicas, Universidad Complutense de Madrid, E-28040 Madrid, Spain
Abstract:A flow injection system has been developed consisting of on-line preconcentration of selenium species in a microcolumn filled with activated alumina, reduction of Se(VI) to Se(IV) and determination by HG-AAS. When 0.01 mol/L HNO3 is used both as carrier and activation reagent for the alumina microcolumn, up to 150 ng of Se(IV) and Se(VI) can be preconcentrated and quantitatively eluted by 500 mgrL of 2 mol/L NH3. The preconcentration factor is 50 when 25 mL of sample is used. The detection limit is about 6 ng/L, the precision is 5% for low concentrations such as 150 ng/L and 3% at high concentrations such as 120 ng/mL. The proposed method is suitable for natural water samples and inorganic Se speciation can be performed by determining Se(IV) and total selenium [Se(VI) is evaluated from the difference].