首页 | 本学科首页   官方微博 | 高级检索  
     检索      


The mechanical pre-stressing in ultrasonic piezotransducers
Authors:Arnold F J  Mühlen S S
Institution:Superior Center of Technology Education, Campinas State University, Limeira, SP, Brazil. arnold@claretianas.com.br
Abstract:Composed piezotransducers submitted to mechanical pre-stressing present shifts on resonance and anti-resonance frequencies. Changes on characteristic parameters of the ceramic and in the coupling between the parts of the transducer can be the causes for this behavior. In applications where the level of pre-stressing is low (up to 50 MPa) the parameters of the ceramic are not altered, therefore, the shifting on frequencies are attributed to coupling between parts. This paper describes a mathematical model to explain this effect based on difference of effective cross-section between transducers parts under pre-stressing. The results show a proportional relation between pre-stressing and effective coupling of the parts.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号