首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of Cadmium(ii), Copper(ii), Mercury(ii), and Lead(ii) in Water Using Stochastic Sensors Based on Graphite and Diamond Paste Modified with 1H-Pyrrole-1-Hexanoic Acid
Authors:Raluca-Ioana Stefan-van Staden  Jacobus Frederick van Staden  Livia Alexandra Gugoasa  Luisa-Roxana Popescu-Mandoc
Institution:1. Laboratory of Electrochemistry and PATLAB, National Institute of Research for Electrochemistry and Condensed Matter, Bucharest, Romania;2. ralucavanstaden@gmail.com;4. The National Research and Development Institute for Industrial Ecology, INCD-ECOIND Bucharest, Bucharest, Romania
Abstract:Graphite/diamond pastes modified with 1H-pyrrole-1-hexanoic acid were used for the design of stochastic sensors. The proposed stochastic sensors were used to determine four heavy metals, copper(II), cadmium(II), mercury(II), and lead(II), in water samples. The sensitivities of the stochastic sensors were high for all four metal ions, and their limits of determination were sufficiently low to enable detection at very minute concentrations; in some cases, unreachable using standard methods of analysis. The recoveries of copper(II), cadmium(II), mercury(II), and lead(II) in water samples were higher than 93.00%, with relative standard deviation values lower than 1.00%. The sensors were used to simultaneously determine copper(II), cadmium(II), mercury(II), and lead(II) in water samples. The obtained results were in agreement with those obtained using standard analytical methods.
Keywords:Cadmium  copper  lead  mercury  pyrrole  stochastic sensor  water analysis
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号