Determination of Tetracycline by Microdroplet Hydrodynamic Adsorptive Voltammetry Using a Multiwalled Carbon Nanotube Paste Rotating Disk Electrode
Authors:Ayesha Sultana  Kazuto Sazawa  Md. Saiful Islam  Kazuharu Sugawara
Affiliation:1. Department of Environmental Biology and Chemistry, Graduate School of Science and Engineering for Research, University of Toyama, Toyama, Japan;2. Maebashi Institute of Technology, Maebashi, Japan
Abstract:A novel and simple method is proposed for the determination of tetracycline by adsorptive voltammetry in a droplet using a carbon nanotube paste rotating disk electrode (CNTP-RDE). An enhanced electrochemical oxidation response of tetracycline was observed in pH 8.2 supporting electrolyte by the addition of a long-chain cationic surfactant, such as benzyldimethyltetradecylammonium chloride (zephiramine). Under the optimized experimental conditions, the calibration curve was linear across a tetracycline concentration range from 1.0?×?10?7 to 2.0?×?10?6 M. The limit of detection and sensitivity were 4.0?×?10?8 M and 0.9358?A M?1, respectively. This method was successfully employed for the determination of tetracycline in milk samples.
Keywords:Adsorptive voltammetry  microdroplet hydrodynamic adsorptive voltammetry  tetracycline  zephiramine