首页 | 本学科首页   官方微博 | 高级检索  
     检索      


An "openable," high-strength gradient set for orthopedic MRI.
Authors:S Crozier  W U Roffmann  K Luescher  C Snape-Jenkinson  L K Forbes  D M Doddrell
Institution:Centre for Magnetic Resonance, University of Queensland, Brisbane, Queensland, 4072, Australia.
Abstract:A novel three-axis gradient set and RF resonator for orthopedic MRI has been designed and constructed. The set is openable and may be wrapped around injured joints. The design methodology used was the minimization of magnetic field spherical harmonics by simulated annealing. Splitting of the longitudinal coil presents the major design challenge to a fully openable gradient set and in order to efficiently design such coils, we have developed a new fast algorithm for determining the magnetic field spherical harmonics generated by an arc of multiturn wire. The algorithm allows a realistic impression of the effect of split longitudinal designs. A prototype set was constructed based on the new designs and tested in a 2-T clinical research system. The set generated 12 mT/m/A with a linear region of 12 cm and a switching time of 100 micros, conforming closely with theoretical predictions. Preliminary images from the set are presented.