Systematic Errors Associated with the CPMG Pulse Sequence and Their Effect on Motional Analysis of Biomolecules
Authors:A Ross  M Czisch  GC King
Institution:aSchool of Biochemistry and Molecular Genetics, University of New South Wales, Sydney, 2052, New South Wales, Australia;bBijvoet Center for Biomolecular Research, Utrecht University, Utrecht, The Netherlands
Abstract:A theoretical approach to calculate the time evolution of magnetization during a CPMG pulse sequence of arbitrary parameter settings is developed and verified by experiment. The analysis reveals that off-resonance effects can cause systematic reductions in measured peak amplitudes that commonly lie in the range 5–25%, reaching 50% in unfavorable circumstances. These errors, which are finely dependent upon frequency offset and CPMG parameter settings, are subsequently transferred into erroneousT2values obtained by curve fitting, where they are reduced or amplified depending upon the magnitude of the relaxation time. Subsequent transfer to Lipari–Szabo model analysis can produce significant errors in derived motional parameters, with τeinternal correlation times being affected somewhat more thanS2order parameters. A hazard of this off-resonance phenomenon is its oscillatory nature, so that strongly affected and unaffected signals can be found at various frequencies within a CPMG spectrum. Methods for the reduction of the systematic error are discussed. Relaxation studies on biomolecules, especially at high field strengths, should take account of potential off-resonance contributions.