Microwave absorption of oxygen measured with a Fabry-Perot spectrometer |
| |
Authors: | R.K.L. Poon |
| |
Affiliation: | Department of Electrical Engineering and Research Laboratory of Electronics, Massachusetts Institute of Technology, Cambridge, MA 02139, U.S.A. |
| |
Abstract: | A Fabry-Perot interferometer in the semiconfocal configuration has been used as a microwave spectrometer by the method of Q-reduction. The Q can be determined with conventional laboratory equipment to within 2% accuracy, corresponding to 0.3 dB/km at 58.82 GHz. Absolute of pure oxygen and dry air at fixed frequencies has been measured as a function of pressure from 0 to 1 atm at room temperature. The measured absorption of dry air can be explained if the overlap of the oxygen lines is properly taken into account. |
| |