Reduction of temperature variation in the atomic absorption graphite furnace |
| |
Authors: | W. Slavin S.A. Myers D.C. Manning |
| |
Affiliation: | The Perkin-Elmer Corporation, Main Avenue, Norwalk, CT 06856 U.S.A. |
| |
Abstract: | The temperature variation that is experienced along the length of the graphite furnace tube of the Massmann design can be reduced by the use of a contoured tube. An analytical model of the steady-state temperature distribution along the graphite tube has been developed and has been shown to agree quite well with experimental data. Steady-state and time-dependent measurements along the length of the graphite tube are reported for different conditions. With conventional tubes at thermal equilibrium, there is a temperature difference exceeding 1000°C between the center and the ends when the center of the tube is at 2500°C. With a contoured tube this temperature gradient has been reduced to 100°C. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |