A comparison of EDI with solvent-free MALDI and LDI for the analysis of organic pigments
Authors:Kudaka Ichiro  Asakawa Daiki  Mori Kunihiko  Hiraoka Kenzo
Institution:Mitsubishi Chemical Group, Science and Technology Research Center, Inc., Kamoshida-1000, Aoba, Yokohama 227-8502, Japan.
Abstract:To evaluate the applicability of EDI to material analysis as a new ionization method, a comparison of EDI with solvent-free matrix-assisted laser desorption ionization (MALDI) and laser desorption ionization (LDI) was made for the analysis of organic pigments, e.g. Pigment Yellow 93, Pigment Yellow 180, and Pigment Green 36, as test samples, which are poorly soluble in standard solvents. In EDI, the samples were prepared in two ways: deposition of suspended samples in appropriate solvents and dried on the substrate, and the direct deposition of the powder samples on the substrate. No matrices were used. Both sample preparation methods gave similar mass spectra. Equally strong signals of M + H](+) and M - H](-) ions were observed with some fragment ions for azo pigments in the respective positive or negative mode of operation. For the powder sample of the phthalocyanine pigment PG36, M(+*) and M + H](+) in the positive mode and M(-*) in the negative mode of operation were observed as major ions. Positive-mode, solvent-free MALDI gave M(+), M + H](+) and M + Na](+) and negative mode gave M - H](-) depending on the sample preparation. As solvent-free MALDI, EDI was also found to be an easy-to-operate, versatile method for the samples as received.
Keywords:EDI  solvent‐free MALDI  LDI  pigment  electrospray