Effects of Injected Volume and Applied Voltage on Column Efficiency in Capillary Electrochromatography with Open Tubular Columns of 10 μm i. d.
Authors:Antonio Luis Crego  Jesús Martínez  María Luisa Marina
Abstract:The effect of some instrumental parameters on column efficiency in open tubular capillary electrochromatography (OTCEC) has been evaluated. First, it was investigated whether band broadening due to the sample injection process is within a tolerable range when an open tubular column (OTC) of about 10 μm i. d. is used. As a result of the small injection profile factor (K2 = 1.3), injected volumes must be sufficiently small (less than 10 pL) to avoid a significant efficiency loss (>5%) when hydrodynamic injection by siphoning is employed. Secondly, the kinetic performance of OTCs in a CEC system was estimated from the variation of the reduced plate height (h) with the reduced linear velocity (ν) which was controlled by the voltage applied. Reasonable agreement was obtained between the theoretical h versus ν curve and the experimental values for a group of polycyclic aromatic hydrocarbons used as test compounds. Values of 0.25 for minimum h at an optimum ν of 16 are estimated, which permit separations with around 400,000 plates per meter to be obtained in less than 5 min. Finally, the possibility of estimating the diffusion coefficients of the solutes in the mobile phase from the plot of the height of a theoretical plate versus electroosmotic flow velocity is shown.
Keywords:Open tubular columns  capillary electrochromatography  column efficiency  injected volume  applied voltage