首页 | 本学科首页   官方微博 | 高级检索  
     


Spectrophotometric nonenzymatic determination of hydrogen peroxide using silver nanoparticles
Authors:A. Yu. Olenin  E. G. Olenina
Affiliation:1.Faculty of Chemistry,Moscow State University,Moscow,Russia;2.Vernadsky Institute of Geochemistry and Analytical Chemistry,Russian Academy of Sciences,Moscow,Russia;3.All-Russia Research Institute of Metrological Service,Moscow,Russia
Abstract:A nonenzymatic method was developed for the detection and quantification of hydrogen peroxide using metallic sols obtained by the reduction of silver compounds with sodium borohydride in the presence of a surface stabilizer. These sols changed color on exposure to aqueous solutions of hydrogen peroxide. The nature of the stabilizer used in sol preparation affects spectral characteristics of the final product formed in the reaction with hydrogen peroxide. In the case of polyvinyl pyrrolidone, the intensity of the surface plasmon resosnance absorption band at 405 nm decreased. In using cetyltrimethylammonium bromide, a signal at 519 nm appeared along with the similar decrease in the absorption band at 408 nm. The band intensity depends on the concentration of hydrogen peroxide. The described phenomena can form a basis for the development of procedures for the qualitative and quantitative determination of hydrogen peroxide in water bodies.
Keywords:
本文献已被 SpringerLink 等数据库收录!