首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A luminescence immunoassay test method for determining benzo[a]pyrene in natural water
Authors:N V Beloglazova  P S Shmelin  A S Novikova  I Yu Goryacheva
Institution:1.Chernyshevskii Saratov State University,Saratov,Russia;2.Tekhnomash,Moscow,Russia
Abstract:A test method is developed for determining benzoa]pyrene in natural water, based on the use of a polyethylene filter (frit) with adsorbed specific antibodies, placed within a transparent column. In passing a test solution, the analyte is adsorbed on the frit similarly to the process implemented in immunoaffinity preconcentration. The added conjugate of a labeled analyte takes the remained vacant binding sites of antibodies. Luminescent semiconductor nanoparticles (quantum dots) CdSe/ZnS, used as labels, enable visual determination under irradiation with UV light. The limit of detection for benzoa]pyrene in water is ~0.5 ng/mL.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏