首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development and validation of a stability-indicating column high-performance liquid chromatographic assay method for determination of nebivolol in tablet formulation
Authors:Kachhadia Pankaj K  Doshi Ashish S  Joshi Hitendra S
Institution:Saurashtra University, Department of Chemistry, Rajkot-360 005, Gujarat, India.
Abstract:A simple, precise, and accurate isocratic reversed-phase (RP) stability-indicating column high-performance liquid chromatographic (HPLC) assay method was developed and validated for determination of nebivolol in solid pharmaceutical dosage forms. Isocratic RP-HPLC separation was achieved on a Phenomenex Luna C8 (2) column (250 mm x 4.6 mm id, 5 microm particle size) using mobile phase composed of acetonitrile-pH 3.5 phosphate buffer (35 + 65, v/v) at a flow rate of 1.0 mL/min, and detection was performed at 280 nm using a photodiode array detector. The drug was subjected to oxidation, hydrolysis, photolysis, and heat to apply stress conditions. The method was validated for specificity, linearity, precision, accuracy, robustness, and solution stability. The method was linear in the drug concentration range of 40-160 microg/mL with a correlation coefficient of 0.9999. The repeatability relative standard deviation (RSD) for 6 samples was 0.69%, and the intermediate precision (RSD) for 6 samples was 1.39%. The accuracy (recovery) was between 98.57 and 99.55%. Degradation products produced as a result of stress studies did not interfere with detection of nebivolol, and the assay can thus be considered stability-indicating.