Performancevalidation of step-isothermal calorimeters
Authors:B A Finnin  M A A O'Neill  S Gaisford  A E Beezer  J Hadgraft  P Sears
Affiliation:(1) Victorian College of Pharmacy, Monash University, Parkville, Victoria , Australia, 3052, 381 Royal Parade;(2) University of London, School of Pharmacy, London , UK, WC1N 1AX, 29-39 Brunswick Square;(3) Thermal Hazard Technology, 1 North House, Bletchley, Buckinghamshire , UK, MK1 1SW, Bond Avenue
Abstract:Isothermal calorimetry is becoming indispensable as a tool for the study of a wide variety of systems. As with all scientific instruments it is essential that robust calibration routines be developed in order to validate the data obtained. Chemical test reactions offer many advantages over (the traditionally used) joule effect heating methods, not least because they have the potential to validate instrument performance (i.e. they can be used to assess all aspects of calorimeter operation). In this work the results of a validation exercise, conducted by Thermal Hazard Technology as part of an installation routine, using the base catalysed hydrolysis of methyl paraben are discussed. In the case described, a systematic misreporting of the reported temperature of a calorimeter was identified, caused by an upgrade to the calorimeter's firmware, a discrepancy which may not have been noted using traditional electrical calibration methods and one which highlights the importance of both manufacturers and end-users adopting chemical test reactions into their test and validation routines.
Keywords:calibration  isothermal calorimetry  methyl paraben  rate constants  validation