首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Results of a joint auger/ESCA round robin sponsored by astm commitree E-42 on surface analysis: Part I. Esca results
Authors:CJ Powell  NE Erickson  TE Madey
Institution:Surface Science Division, National Bureau of Standards, Washington. D.C. 20234 U.S.A.
Abstract:We report results of a round robin involving binding-energy (BE) and relative-intensity measurements on high-purity samples of gold and copper by X-rayWe have conducted an extensive round robin involving ESCA or XPS measurements on high-purity samples of Ni, Au, and Cu. The Ni sample was used to estabThis round robin had three principal objectives. First, it was intended to assess the overall accuracy of BE and intensity measurements in XPS measuremSecond, it was intended to establish an empirical intensity calibration of one type of XPS instrument versus another from the reported intensity ratiosThird, it was desired to determine the peak position and peak shape for the carbon 1s photoelectron line originating for “adventitious” carbonaWe have made an extensive analysis of trends in the BE and intensity data. We have computed deviations of individual BE measurements from the median vaThe two types of plots just described based on BE and intensity data can be regarded as “response functions” that describe the performance of indivWe have invoked Occam's Razor to distinguish instruments with response functions which vary monotonically with electron energy from those instruments wThe reported spreads in BE and relative-intensity data reported here indicate that improved calibration and operating procedures are required for XPS m
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏