首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Improved in vivo photoacoustic microscopy based on a virtual-detector concept
Authors:Li Meng-Lin  Zhang Hao E  Maslov Konstantin  Stoica George  Wang Lihong V
Institution:Optical Imaging Laboratory, Department of Biomedical Engineering, Texas A&M University, 3120 TAMU, College Station, Texas 77843-3120, USA.
Abstract:Recently an in vivo high-resolution backward-mode photoacoustic microscope was developed that shows potential for applications in dermatology and related cancer research. However, the limited depth of focus of the large-numerical-aperture (NA) ultrasonic lens employed in this system causes the image quality to deteriorate significantly in the out-of-focus region. To solve this problem, we devised and explored, for the first time to our knowledge, a virtual-detector-based synthetic-aperture focusing technique, combined with coherence weighting, for photoacoustic microscopy with such a large-NA transducer. Images of phantoms show that the proposed technique improves the -6 dB lateral resolution from 49-379 to 46-53 microm and increases the signal-to-noise ratio by up to 29 dB, depending on the distance from the ultrasonic focal point. In vivo experiments show that the technique also provides a clearer representation of the vascular distribution in the rat's scalp.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏