首页 | 本学科首页   官方微博 | 高级检索  
     


Electrolytic conductivity: a reference system for aqueous solutions
Authors:Ferrara E  Durbiano F  Capra P P  Marullo Reedtz G
Affiliation:Istituto Elettrotecnico Nazionale Galileo Ferraris, Strada delle Cacce, 91, I-10135 Torino, Italy. ferrara@ien.it
Abstract:To establish a system for the measurement of electrolytic conductivity in aqueous solutions with traceability to the SI, a cell has been developed with a removable central section providing an accurate determination of the geometrical constant. As the concentration unit mass/mass, based and referred to the amount of substance in the SI, is recommended by international organisations, 0.1 molal KCl samples have been prepared and analysed. Results over a temperature range from 20 degrees C to 35 degrees C are presented. The main sources of error have been estimated and relative uncertainty has been assessed to remain at the level of 0.6%. Satisfactory agreement is observed when the results are compared with the reference values available from literature.
Keywords:
本文献已被 PubMed 等数据库收录!