The determination of silicon in deionized process water by graphite furnace AAS
Authors:F. Fehse
Affiliation:Varian GmbH Alsfelderstrasse 6 Postfach 11 11 54 D-6100 DarmstadtFederal Republic of Germany
Abstract:A method for automated in situ preconcentration in graphite furnace AAS is described. The method permits the determination of silicon in deionized process water at the 0.9 μg 1?1 level with a relative standard deviation of 5%.