首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Liquid membrane potentiometric sensor for determination of Fe3+ ion
Authors:Hassan Ali Zamani  Farnoush Faridbod
Institution:1. Department of Applied Chemistry, Mashhad Branch, Islamic Azad University Mashhad, Mashhad, Iran
Abstract:Solution studies showed a selective interaction between the new synthesized ionophore, 2-(thiophen-2-yl)methyleneamino]isoindoline-1,3-dione (TMID) and Fe(III) ion. Therefore, TMID was used as an iron selective ion-carrier in a plasticized liquid membrane sensor. The linear response range of the proposed electrode was 1.0 × 10?6–1.0 × 10?2 M. The Nernstian slope of 20.1 ± 0.3 mV/decade, and a detection limit of 5 × 10?7 M was obtained. The sensor could be used in the pH range of 2.3–4.8, and the response time was about 10 s. The lifetime of the electrode was at least 7 weeks. The sensor accuracy was investigated in two ways: (i) with the potentiometric titration of a Fe3+ solution with EDTA and (ii) with Fe(III) monitoring in some cationic mixtures. Finally, the newly fabricated electrode was effectively applied as an indicator electrode for the direct Fe3+ determination in real samples.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏