Calibration and applications of polarized neutron thermometry at milli- and microkelvin temperatures
Authors:T. A. Jyrkkiö  M. T. Huiku  K. N. Clausen  K. Siemensmeyer  K. Kakurai  M. Steiner
Affiliation:(1) Low Temperature Laboratory, Helsinki University of Technology, SF-02150 Espoo, Finland;(2) Present address: Physics Department, Risø National Laboratory, P.O. Box 49, DK-4000 Roskilde, Denmark;(3) Hahn-Meitner-Institut, Glienicker Strasse 100, D-1000 Berlin, Germany
Abstract:We have utilized the polarized neutron scattering technique for the determination of the nuclear temperature in copper from 20 mK down to 100 mgrK. Using a cooling facility where two adiabatic demagnetization stages work in series we have calibrated the flipping ratio of neutrons scattered from the (200) Bragg-peak against the nuclear polarization over the whole polarization range. The observed calibration curve deviates considerably from the curve expected for an extinction-free sample. The polarized neutron thermometer is discussed and applied in measurements of the spin-lattice relaxation in copper.