Integrated system for on-line gas and liquid chromatography with a single mass spectrometric detector for the automated analysis of environmental samples
Authors:J Slobodník  A C Hogenboom  A J H Louter  U A Th Brinkman
Institution:

Free University, Department of Analytical Chemistry, De Boelelaan 1083, 1081 HV, Amsterdam, Netherlands

Abstract:An integrated system has been developed which combines liquid (LC) and gas (GC) chromatographic separation with a single mass spectrometer (MS). On-line solid-phase extraction (SPE) of 10–200 ml aqueous samples on a short (10 × 2.0 mm I.D.) precolumn packed with a styrene-divinylbenzene copolymer is used for analyte enrichment. The trace-enrichment procedure was automated by means of a PROSPEKT cartridge-exchange/solvent-selection/valve-switching unit. After sample loading, the precolumn is eluted on-line in two subsequent runs, first onto the GC-MS system and, next, onto the LC-MS system using a particle beam (PB) interface. Prior to entering the PB-MS, the LC eluent passes through the flow cell of a UV diode-array detector (DAD). Both GC-MS and LC-PB-MS generate classical electron ionisation (EI) and chemical ionisation (CI) spectra which are useful for the identification of low- and sub-μg/l concentrations of environmental pollutants covering a wide polarity and volatility range. The LC-DAD data provide additional means for quantitation and yield complementary spectral information. All three detection systems (GC-MS, LC-DAD, LC-PB-MS) and the trace-enrichment procedure are fully automated and controlled from the keyboard of the central computer. With such a ‘MULTIANALYSIS’ system GC-MS, LC-DAD and LC-MS data of the same sample can be obtained within 3 h. The system was optimised with nine chlorinated pesticides in drinking water as test mixture. With 100-ml samples detection limits in GC-MS were 0.0005?0.03 μg/l, and in LC-PB-MS 0.5–7 μg/l, both in the full-scan (EI) mode. Negative chemical ionisation (NCI) with methane as reagent gas improved the sensitivity of six halogenated compounds 3- to 30-fold and provided relevant information for structural elucidation of unknown compounds in real-world samples. LC-DAD detection limits varied from 0.01 to 0.05 μg/l. Relative standard deviations (R.S.D.) of retention times were less than 0.2% in all systems, R.S.D.s of peak areas were 5–15% for GC-MS and LC-PB-MS and less than 5% for LC-DAD. The ‘MULTIANALYSIS’ system was used to analyse surface water samples and river sediment extracts; several pollutants were detected and identified.
Keywords:Environmental analysis  Water analysis  Automation  Liquid chromatography-mass spectrometry  Gas chromatography-mass spectrometry Pesticides
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号