Optimization of a conversion electron Mössbauer spectroscopy gas flow He/CH4 proportional counter
Authors:Darko Hanzel  P. Griesbach  W. Meisel  P. Gütlich
Affiliation:(1) J. Stefan Institute, University of Ljubljana, YU-61111 Ljubljana, Slovenia, Yugoslavia;(2) Institute of Inorganic and Analytical Chemistry, University of Mainz, D-6500 Mainz, Germany
Abstract:A new detector for CEMS has been built and optimized with respect to the statistical quality of spectra obtained. The optimization has been performed by measuring Mößbauer and pulse height spectra at in- and off-resonance. Single channel analyzer settings were calculated by a new optimization routine. A comparison of different detector designs has been performed using the statistical utility rate of spectra obtained from a stainless steel foil. A procedure for determining optimal operating parameters for ICEMS gas flow proportional counters is proposed.