Simultaneous Electroanalysis of Hypochlorite and H2O2: Use of I−/I2 as a Probing Potential Buffer
Abstract:Sodium hypochlorite (NaClO) and hydrogen peroxide (H2O2) have been simultaneously analyzed, for the first time, using a simple and rapid potentiometric method. The present method shows a high sensitivity, selectivity and satisfactory reproducibility. Pt electrode was used as an indicator electrode and the I2/I? redox couple was used as a probing potential buffer. The large difference in the rates of the oxidation of I? by the two oxidizing agents, that is, the oxidation of I? by NaClO is by several orders of magnitude faster than that by H2O2, enabled the selective analysis of these two species. Based on such a large difference in the rates, a momentary potential response corresponding to the oxidation of I? by NaClO and another quite slow one by H2O2 could be obtained. Factors affecting the selectivity as well as the sensitivity, such as the concentrations of molybdate (used as catalyst for the oxidation of I? by H2O2), H+, I2, and I? in the potential buffer were examined. The expected Nernstian responses were obtained over a considerable range of the concentrations of the two oxidizing agents with slopes of 30.5 and 29.9 mV for NaClO and H2O2, respectively (in a close agreement with the theoretical value, that is, 29.6 mV) and with a detection limit in the submicromolar range (0.2 μM).
Keywords:Hypochlorite  Hydrogen peroxide  Electroanalysis  Potentiometry  Water quality