Determination of Selenium with a Poly(1,8‐diamino‐naphthalene)‐Modified Electrode
Abstract:The performance of a poly(1,8‐diaminonaphthalene)‐modified electrode for the determination of the Se(IV) ion in an aqueous medium was investigated with anodic stripping voltammetry without the pretreating of the sample. The experimental parameters for the analysis of Se(IV) were optimized and the characteristics of this polymer‐modified electrode were investigated by using cyclic voltammetry. The Se(IV) ions were chemically deposited onto the surface of the pDAN‐Au electrode in an acidic medium. The detection limit employing the anodic stripping differential pulse voltammetry was 9.0×10?9 M for Se(IV) with 4.4 % of RSD. Satisfactory result for the determination of Se(IV) was acquired employing a certified standard urine reference material, SRM's 2670 (trace element in urine) with 4.1 ppb of SD.
Keywords:Poly(1  8‐diaminonaphthalene)  Modified electrode  Selenium analysis  Stripping differential pulse voltammetry