The heat exchange between the horizontal furnace of a differential thermal analyzer (DTA) and the sample is analyzed with the aim of understanding the parameters governing the thermal signal. The resistance due to radiation and conduction through the gas has been calculated and compared to the experimental values of the thermal-lag between the sample and furnace and apparatus sensitivity. The overall evolution of these parameters with the temperature and their relative values are well understood by considering the temperature differences that arise between the sample and holder. Two RC thermal models are used for describing the apparatus performance at different temperature ranges. Finally, the possibility of improving the signal quality through the control of the leak resistances is stressed.