首页 | 本学科首页   官方微博 | 高级检索  
     


Quantitative measurements on wetted thin layer chromatography plates using a charge coupled device camera
Authors:Lancaster Michael  Goodall David M  Bergström Edmund T  McCrossen Sean  Myers Peter
Affiliation:Department of Chemistry, University of York, York YOI0 5DD, UK.
Abstract:This paper presents the first study of imaging of spots on thin-layer chromatographic plates whilst still wet with solvent. Imaging and quantification of Sudan II after development with dichloromethane was carried out in both reflectance and transmission modes, using a charge coupled device (CCD) camera. The relationship between peak area and sample loading was established at low sample loading, and found to be linear over an order of magnitude for both wet and dry modes with r2-values > 0.99. All data processing was carried out using the Beer-Lambert equation. Curvature at high loadings in the plots of integrated absorbance as a function of sample loading was accounted for using an empirical expression designed for use with the Kubelka-Munk treatment and apparent absorbance of the stationary phase due to scattering. Results are consistent with an effective pathlength significantly longer than the thickness of the sorbent layer. The limit of detection on a dry plate (0.5 ng) was found to be lower than on a wetted plate (2 ng). Precision was found to be 1-4% RSD intra-plate and 8-14% RSD inter-plate. Results are compared with quantification of the same analyte on dried plates.
Keywords:
本文献已被 PubMed 等数据库收录!