a China Institute of Atomic Energy, P.O. Box 275(96), Beijing 102413, China
* Chinese Nuclear Society, P.O. Box 2125, Beijing 100045, China
Abstract:
Two techniques have been developed to investigate the characteristic of the sensitivity of CR-39 (polyallyldiglycol carbonate) track detector using Am---Be neutron source. The firstis to measure volume track density created by fast neutrons in CR-39 detectors. The second is to measure areal asymptotic track density on the surface of CR-39 sheets. Two types of CR-39—SY-2 CR-39 made in China and another made in USA were calibrated using the techniques. It shows that the two types of CR-39 detectors have nearly the same sensitivity to detect fast neutrons. The techniques developed in this work can be applied to calibration of other track detectors.