首页 | 本学科首页   官方微博 | 高级检索  
     


Wide-bore monolithic column for electrochromatography
Authors:Yuan Ruijuan  Ding Guosheng  Guo Yugao  Liu Danning  Bao James J
Affiliation:School of Pharmaceutical Science and Technology, Tianjin University, Tianjin, P. R. China.
Abstract:A new wide-bore electrophoresis (WE) system adopting an inner cooling device was set up to perform electrochromatography. In this system, a quartz tube of 1.2 mm inner diameter was used as the separation channel. The Joule heat generated during electrophoresis was removed timely through the outer surface of the quartz tube and a cooling capillary inserted into the quartz tube. A proper coolant passed through the cooling capillary to further improve the cooling efficiency. In the primary research, a polyacrylamide monolithic column was successfully prepared in this quartz tube. Then it was evaluated in the electrochromatographic mode. An electric field strength as high as 625 V/cm can be applied to this system without obvious deviation of the current from the linear curve of the Ohm plot. Sample volume as high as 1 microL was injected into the WE system and reasonable efficiency was obtained for separation of the test compounds.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号