首页 | 本学科首页   官方微博 | 高级检索  
     


Multifrequency Probe for Pulsed EPR and ENDOR Spectroscopy
Authors:Nikolai I. Avdievich  Gary J. Gerfen
Affiliation:Department of Physiology and Biophysics, Albert Einstein College of Medicine, 1300 Morris Park Avenue, Bronx, New York 10461, USA. navdievi@aecom.yu.edu
Abstract:The design, construction, and performance of a multifrequency pulsed EPR and ENDOR probe for use at cryogenic temperatures are described. Interchangeable resonators based on a folded strip line design allow variation of the resonance frequency over a range of 5-11 GHz. Variable coupling to the resonator is achieved capacitively via a simple mechanical adjustment which is thermally and mechanically stable. The entire assembly is robust and easily fabricated. Common methods of analyzing the resonator parameters such as the Q-factor and coupling coefficient are discussed quantitatively. Probe performance data and multifrequency pulsed ENDOR spectra are presented.
Keywords:Abbreviations: EPR instrumentationAbbreviations: pulsed ENDORAbbreviations: ESEEMAbbreviations: resonatorAbbreviations: coupling measurement