Amperometric urea biosensor based on urease and electropolymerized toluidine blue dye as a pH-sensitive redox probe |
| |
Authors: | Vostiar Igor Tkac Jan Sturdik Ernest Gemeiner Peter |
| |
Affiliation: | Department of Biochemical Technology, Faculty of Chemical and Food Technology, Slovak University of Technology, Radlinskeho 9, SK-812 37 Bratislava, Slovak Republic. vostiar@chelin.chtf.stuba.sk |
| |
Abstract: | The electropolymerized toluidine blue film deposited on the glassy carbon electrode show amperometrically detectable pH sensitivity. This feature of polytoluidine blue (PTOB) film was used for a construction of an amperometric urea biosensor. We have observed a linear shift of the formal redox potential with increasing pH value between 4 and 8 giving the slope of 81 mV(Delta) pH(-1). Polytoluidine blue film has had a significantly increased stability and higher electrochemical activity compared to the adsorbed monomeric dye. The polytoluidine blue urea biosensor has been operating at a working potential of -200 mV vs. SCE. The sensitivity of the biosensor was 980 nA mM(-1) cm(-2). The biosensor showed linearity in concentration range up to 0.8 mM with the detection limit of 0.02 mM (S/N=3). |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|