首页 | 本学科首页   官方微博 | 高级检索  
     检索      


In situ measurements of subsurface contaminants with a multi-channel laser-induced fluorescence system.
Authors:Jane Wu Pepper  Andrew O Wright  Jonathan E Kenny
Institution:Department of Chemistry and Center for Field Analytical Studies and Technology, Tufts University, Medford, MA 02155, USA. jpepper@srubiosystems.com
Abstract:A new multi-channel laser-induced fluorescence (LIF) probe with novel optical fiber probe geometry has been designed and integrated into a cone penetrometer testing (CPT) system for in situ contamination detection. The system is capable of collecting excitation and emission matrices (EEMs) of subsurface contaminants as a function of depth in seconds. Compared to our previous multi-channel LIF-CPT system, the new system is faster and more compact, with reduced probe size and sampling area. This article describes the first field demonstration of the system at Hanscom Air Force Base, Massachusetts. One contaminated site within the base was characterized through in situ measurements of 26 LIF-CPT pushes. To validate the LIF results, core samples taken at five locations were analyzed by both on-site LIF measurements and by off-site laboratory analyses with EPA methods. The comparison of the LIF and laboratory results is presented, along with the results of the in situ measurements.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏